DISPlay:TEXT:CLEar
Syntax
DISPlay[:WINDow]:TEXT:CLEar
Description
This command clears the text message displayed on the instrument's front-panel display (see
DISPlay:TEXT command).
Remarks
l Clearing the text message does not override the display state (see DISPlay:STATe command); if the dis-
play was disabled prior to clearing the text message, the display will remain disabled.
l If the text message was displayed by overriding a displayed measurement or math function, that func-
tion will reappear when the text message is cleared.
l The front-panel display is automatically cleared after a Factory Reset (*RST command), or after an
Instrument Preset (SYSTem:PRESet command).
Examples
The following example clears the text message on the top line (the default) of the front panel.
DISP:TEXT:CLE
See Also
DISPlay:TEXT