STATus:QUEStionable:ENABle
Syntax
STATus:QUEStionable:ENABle <enable>
STATus:QUEStionable:ENABle?
Description
This command enables bits in the enable register for the Questionable Data Register group. The selected
bits are then reported to the Status Byte.
For more information on the SCPI Status System for the Keysight, see the Status Sys-
tem Diagram.
Parameters
Name Type Range of Values Default Value
<enable> Numeric A decimal value which corresponds to the
binary-weighted sum of the bits in the
register (see table below).
This is a required parameter
Remarks
l The following table lists the bit definitions for the Questionable Data Register.
Bit Num-
ber
Bit Name
Decimal
Value
Definition
0
Voltage Overload
1 The input termination protection relay is open.
1 Not Used 2 (Reserved for future use)
2 Time 4 A timeout or measurement overflow occurred during a time inter-
val, rise/fall time, duty cycle, pulse width or single-period meas-
urement.
3 Not Used 8 (Reserved for future use)
4 Not Used 16 (Reserved for future use)
5 Freq 32 A timeout or measurement overflow occurred during a frequency,
frequency ratio, average period, PRF, or PRI measurement..
6 Phase 64 A timeout or measurement overflow occurred during a phase meas-
urement.
7 Not Used 128 (Reserved for future use)
8 CAL Corrupt 256 At least one calibration constant is corrupt.
STATus:QUEStionable:ENABle
392 Keysight 53220A/53230A Programmer's Reference