CALCulate1:AVERage:COUNt:CURRent?
Syntax
CALCulate[1]:AVERage:COUNt:CURRent?
Description
This query returns the number of measurements used to compute the current statistics values since the
last time statistics were cleared.
Remarks
l You can read the statistical values at any time.
l The instrument clears the statistical values when CALCulate1:AVERage:STATe ON is sent, when the
measurement function is changed, when the INITiate:IMMediate, MEASure? or READ? commands are
executed, when the CALCulate1:AVERage:CLEar command is executed, after a Factory Reset (*RST
command) or after an Instrument Preset (SYSTem:PRESet command).
Return Format
The query returns the number of measurements used to compute the current statistics values in the form
+100.
Examples
The following example returns the number of measurements used to compute the current statistics val-
ues.
CALC:AVER:COUN:CURR?
Typical Response: +73
See Also
CALCulate1:AVERage:CLEar
CALCulate1:AVERage:STATe
CALCulate1:STATe
Keysight 53220A/53230A Programmer's Reference 71
CALCulate1:AVERage:COUNt:CURRent?