EasyManua.ls Logo

Keysight 53230A - Sense:tinterval:gate:polarity

Keysight 53230A
518 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SENSe:TINTerval:GATE:POLarity
Syntax
[SENSe:]TINTerval:GATE:POLarity {POSitive | NEGative}
[SENSe:]TINTerval:GATE:POLarity?
Description
This command selects the polarity for the Gate In/Out BNC when externally gating of time interval, rise/-
fall time, duty cycle, pulse width, phase and single-period measurements is selected (SENSe:TINTer-
val:GATE:SOURce EXTernal selected).
l POSitive starts the measurement on a positive edge on the Gate In/Out BNC.
l NEGative starts the measurement on a negative edge on the Gate In/Out BNC.
Remarks
l The instrument selects negative polarity after a Factory Reset (*RST command) or an Instrument Preset
(SYSTem:PRESet command).
Return Format
The query returns the gate open slope: POS or NEG.
Examples
The following example makes a rise time measurement on channel 1 using external gating. The meas-
urement starts on a negative edge on the Gate In/Out BNC:
CONF:RTIM 20,80
TINT:GATE:SOUR EXT
TINT:GATE:POL NEG
READ?
Typical Response: +1.21513398300000E-006
See Also
SENSe:TINTerval:GATE:SOURce
SENSe:TINTerval:GATE:POLarity
370 Keysight 53220A/53230A Programmer's Reference

Table of Contents

Other manuals for Keysight 53230A

Related product manuals