SENSe:GATE:STOP:HOLDoff:TIME
Syntax
[SENSe:]GATE:STOP:HOLDoff:TIME {<time> | MINimum | MAXimum | INFinity | DEFault}
[SENSe:]GATE:STOP:HOLDoff:TIME? [{MINimum | MAXimum | DEFault}]
Description
This command selects the gate close holdoff time when SENSe:GATE:STOP:HOLDoff:SOURce is set to
TIME.
Parameters
Name Type Range of Values Default Value
<time> Numeric Time Interval measurements:
0.0 to 1000 sec with
10 usec resolution
or +9.9E+37 (INFinity) (Keysight 53220A)
0.0, or 30 nsec to 1000 sec with
10 nsec resolution
or +9.9E+37 (INFinity) (Keysight 53230A)
MIN=0.0 sec
MAX=1000 sec
Frequency, frequency ratio, average period,
PRF, PRI and totalize measurements:
100 usec to 1000 sec with
10 usec resolution (Keysight 53220A)
1 usec to 1000 sec with
1 usec resolution (Keysight 53230A)
MIN=100 usec (Keysight 53220A)
MIN = 1 usec (Keysight 53230A)
MAX=1000 sec
0.0 sec
.1 sec
Remarks
l INFinity or +9.9E+37 is accepted only for the totalize measurement function. When selected, the gate
remains open until an ABORt or *RST command or a device clear is received.
l The instrument sets the gate close holdoff time to 0.0 (no holdoff) after a Factory Reset (*RST com-
mand) or an Instrument Preset (SYSTem:PRESet command).
SENSe:GATE:STOP:HOLDoff:TIME
358 Keysight 53220A/53230A Programmer's Reference