Return Format
The query returns the gate close holdoff time in the form +1.00000000000000E-007. For a infinite gate
close holdoff time, the query returns 9.9E+37.
Examples
The following example makes a frequency measurement on channel 1. The gate opens .1 sec after a neg-
ative edge on the Gate In/Out BNC, and, after .25 seconds, closes on the next positive edge on the BNC.
CONF:FREQ 1.2E6
FREQ:GATE:SOUR ADV
GATE:STAR:SOUR EXT
GATE:STAR:SLOP NEG
GATE:STAR:DEL:SOUR TIME
GATE:STAR:DEL:TIME .1
GATE:STOP:SOUR EXT
GATE:STOP:SLOP POS
GATE:STOP:HOLD:SOUR TIME
GATE:STOP:HOLD:TIME .25
GATE:EXT:SOUR BNC
READ?
Typical Response: +1.21513398300000E+006
See Also
SENSe:GATE:STOP:HOLDoff:SOURce
SENSe:FREQuency:GATE:TIME
SENSe:TOTalize:GATE:TIME
Keysight 53220A/53230A Programmer's Reference 359
SENSe:GATE:STOP:HOLDoff:TIME