CALCulate2:TRANsform:HISTogram:CLEar
Syntax
CALCulate2:TRANsform:HISTogram:CLEar[:IMMediate]
Description
This command clears the histogram data. If automatic scaling is enabled (CALCu-
late2:TRANsform:HISTogram:RANGe:AUTO ON set), the automatic scaling will be restarted.
Remarks
l The instrument clears the histogram data when the CALCulate2:TRANsform:HISTogram:POINts,
CALCulate2:TRANsform:HISTogram:RANGe:AUTO, CALCu-
late2:TRANsform:HISTogram:RANGe:AUTO:COUNt, CALCu-
late2:TRANsform:HISTogram:RANGe:LOWer, CALCulate2:TRANsform:HISTogram:RANGe:UPPer
commands are sent, when CALCulate2:TRANsform:HISTogram:STATe ON is sent, when the meas-
urement function is changed, when the INITiate:IMMediate, MEASure? or READ? commands are
executed, when the CALCulate2:TRANsform:HISTogram:CLEar command is executed, after a Factory
Reset (*RST command) or after an Instrument Preset (SYSTem:PRESet command).
Examples
The following example clears the histogram data.
CALC2:TEAN:HIST:CLE
See Also
CALCulate2:TRANsform:HISTogram:STATe
Keysight 53220A/53230A Programmer's Reference 83
CALCulate2:TRANsform:HISTogram:CLEar