LXI:MDNS:SNAMe:DESired
Syntax
LXI:MDNS:SNAMe:DESired "<name>"
LXI:MDNS:SNAMe:DESired?
Description
This command sets the desired mDNS service name.
Parameters
Name Type Range of Values Default Value
<name> Quoted String A string of up to 63 characters This is a required parameter
Remarks
l The desired mDNS servicename is stored in non-volatile memory, and does not change when power has
been off, after a Factory Reset (*RST command), or after an Instrument Preset (SYSTem:PRESet com-
mand).
l The desired mDNS service name is set to "Keysight 53220A Universal Frequency Counter/Timer - <serial_
number>" (Keysight 53220A) or "Keysight 53230A Universal Frequency Counter/Timer - <serial_
number>" (Keysight 53230A), where <serial number> is the instrument's serial number, when the instru-
ment is shipped from the factory or after a SYSTem:SECurity:IMMediate command.
Return Format
The query returns the desired mDNS service name as an ASCII string enclosed in double quotes.
Examples
The following example defines a desired mDNS service name.
LXI:MDNS:SNAM:DES "LAB1-CNTR"
See Also
LXI:MDNS:ENABle
LXI:MDNS:SNAMe:RESolved?
Keysight 53220A/53230A Programmer's Reference 233
LXI:MDNS:SNAMe:DESired