SYSTem:COMMunicate:LAN:CONTrol?
Syntax
SYSTem:COMMunicate:LAN:CONTrol?
Description
This command reads the initial Control connection port number for Sockets communications. The initial
connection is used to send and receive commands, queries, and query responses.
Remarks
l The Control socket connection is used to send a Device Clear to the instrument or to detect pending Ser-
vice Request (SRQ) events.
Return Format
The query returns the Control connection port number. 0 is returned if the interface does not support a
Socket Control connection.
Example
The following example returns the Control connection port number.
SYST:COMM:LAN:CONT?
Typical Response: 5000
See Also
SYSTem:COMMunicate:LAN:CONTrol?
404 Keysight 53220A/53230A Programmer's Reference