MEASure:TOTalize:TIMed?
Syntax
MEASure:TOTalize:TIMed? [{<gate_time> | MINimum | MAXimum | DEFault}] [,<channel>]
Description
This query sets all measurement parameters and trigger parameters to default values for timed totalize
measurements, then immediately triggers a measurement. The results are sent directly to the instrument
output buffer.
Parameters
Name Type Range of Values Default Value
<gate_time> Numeric 1 usec to 1000 sec
MIN = 1 usec
MAX = 1000 sec
.1 sec
<channel> Expression (@1) | (@2) (@1)
Remarks
l The input channel is configured to use the currently programmed input threshold voltage and slope
(see INPut:LEVel and INPut:SLOPe) with auto-leveling disabled.
l The gate source (see SENSe:TOTalize:GATE:SOURce) is set to TIME and the gate time (see SENSe:TOTal-
ize:GATE:TIME) is set to the <gate_time> value.
Return Format
The query returns a single measurement. The format of the measurement is controlled by the FORMat
Subsystem commands.
Examples
The following example configures the instrument for timed totalize measurements on channel 1 using
default parameters, makes one measurement, and then transfers the measurement to the instrument out-
put buffer.
MEAS:TOT:TIM?
Typical Response: +5.10500000000000E+003
The following example configures the instrument for 1 second timed totalize measurements on channel 2,
makes one measurement, and then transfers the measurement to the instrument output buffer.
Keysight 53220A/53230A Programmer's Reference 277
MEASure:TOTalize:TIMed?