INPut3:BURSt:LEVel
Syntax
INPut3:BURSt:LEVel {<dB> | MINimum | MAXimum | DEFault}
INPut3:BURSt:LEVel? [{MINimum | MAXimum | DEFault}]
Description
This command sets the threshold in terms of dB below the peak amplitude of the envelope of a pulsed
input signal. This threshold is used for all pulse measurements. This setting can be used to compensate
for known overshoot in the pulsed input signal.
The threshold values are approximate. These values are not calibrated or specified.
Parameters
Name Type Range of Values Default Value
<value> Numeric -6 or -12 dB -6
Remarks
l Available only on the Keysight 53230A. Requires option 150.
l The instrument sets the pulse envelope threshold to -6 dB after a Factory Reset (*RST command) or an
Instrument Preset (SYSTem:PRESet command).
Keysight 53220A/53230A Programmer's Reference 225
INPut3:BURSt:LEVel