MEASure:FREQuency:BURSt?
Syntax
MEASure:FREQuency:BURSt? [<channel>]
Description
This query sets all measurement parameters and trigger parameters to default values to measure the car-
rier frequency of a burst (pulse frequency), then immediately triggers a measurement. The results are sent
directly to the instrument output buffer.
Parameters
Name Type Range of Values Default Value
<channel> Expression (@3) (@3)
Remarks
l Available only on the Keysight 53230A. Requires option 150.
l CONFigure and MEASure disable narrow burst mode. Because of this, MEASure:FREQuency:BURSt?
will return 9.91E37 (Not a Number) for bursts shorter than 10 microseconds. For such bursts, you
should use CONFigure:FREQuency:BURSt, SENSe:FREQuency:BURSt:GATE:NARRow ON, and READ?
instead of MEASure:FREQuency:BURSt?.
Return Format
The query returns a single measurement. The format of the measurement is controlled by the FORMat
Subsystem commands.
Examples
The following example configures the instrument for burst frequency measurements on channel 3 using
default parameters, makes one measurement, and then transfers the measurement to the instrument out-
put buffer. The burst must have a duration of at least 10 microseconds.
MEAS:FREQ:BURS?
Typical Response: +4.30013398300000E+009
See Also
CONFigure?
CONFigure:FREQuency:BURSt
Keysight 53220A/53230A Programmer's Reference 245
MEASure:FREQuency:BURSt?