SENSe:ROSCillator:EXTernal:CHECk
Syntax
[SENSe:]ROSCillator:EXTernal:CHECk ONCE
Description
This command checks whether a valid external reference signal is present on the Ext Ref In BNC. An error
is generated if valid signal is not present.
Remarks
l SENSe:ROSCillator:SOURce EXTernal and SENSe:ROSCillator:SOURce:AUTO OFF must be set, or a set-
tings conflict error will be generated.
Examples
The following example checks for a valid reference signal on the Ext Ref In BNC:
ROSC:SOUR EXT
ROSC:EXT:CHEC ONCE
See Also
SENSe:ROSCillator:SOURce
SENSe:ROSCillator:SOURce:AUTO
SENSe:ROSCillator:EXTernal:CHECk
360 Keysight 53220A/53230A Programmer's Reference