CONFigure?
Syntax
CONFigure?
Description
This command returns the function configured by the last CONFigure or MEASure? command.
Remarks
l At power-on, this command generates a settings conflict error until the first CONFigure or MEASure?
command is sent.
l If the instrument state has been changed via commands other than CONFigure or MEASure?, the instru-
ment will not track these changes, and the query response will not reflect these changes.
Return Format
The command returns a series of comma-separated fields indicating the present function, expected value
and resolution or reference level(s), and channel(s). The short form of the function name is always
returned (e.g., "FREQ", "PDUT", etc.).
Examples
The following example returns the present configuration of the meter (the quotes are also returned).
CONF?
Typical Response: "FREQ +1.20000000000000E+006,+1.00000000000000E-006,(@2)"
See Also
CONFigure
CONFigure?
156 Keysight 53220A/53230A Programmer's Reference