AUToscale
Syntax
AUToscale
Description
This command causes the instrument to perform an autoscale operation on the input signal. This will
adjust the input threshold, input range, and internal measurement hardware settings based on the input
signal levels and frequency.
Remarks
l You should select the measurement function and active input channel(s) prior to sending the AUTo-
scale command.
Example
The following example configures a frequency measurement, performs an autoscale, and makes a meas-
urement on channel 1.
CONF:FREQ
AUT
READ?
Typical Response: +1.21513398300000E+006