CALCulate1:SMOothing:STATe
Syntax
CALCulate[1]:SMOothing[:STATe] {OFF | ON}
CALCulate[1]:SMOothing[:STATe]?
Description
This command enables or disables a smoothing (moving average) filter on the measurements.
Remarks
l The CALCulate1 subsystem must also be enabled by CALCulate1:STATe for smoothing (moving aver-
age) to be performed.
l The smoothing (moving average) filter is disabled for continuous totalize since this computation
requires more than one measurement to be useful.
l The CALCulate subystem and the individual calculate functions are disabled for timestamp meas-
urements.
l The instrument disables the smoothing (moving average) filter after a Factory Reset (*RST command) or
an Instrument Preset (SYSTem:PRESet command).
Return Format
The query returns the current smoothing (moving average) filter state: 0 (OFF) or 1 (ON).
Examples
The following example enables a 50-point moving average filter on 1000 frequency measurements.
CONF:FREQ 1.0E6,.001
SAMP:COUN 1000
CALC:SMO:RESP MED
CALC:SMO:STAT ON
CALC:STAT ON
READ?
Typical Response: +9.99383828200000E+06,+9.99352315400000E+06,...
(1000 measurements)
See Also
CALCulate1:SMOothing:RESPonse
CALCulate1:STATe
CALCulate1:SMOothing:STATe
38 Keysight 53220A/53230A Programmer's Reference