FORMat:PHASe
Syntax
FORMat:PHASe {AUTO | POSitive | CENTered}
FORMat:PHASe?
Description
This command selects whether phase measurements are displayed in the fixed range of ±18 or to
360°, or whether the instrument will automatically switch between these ranges based on the meas-
urement history.
Remarks
l AUTO is the algorithm used in the Agilent 53131A/132A for phase measurements.
l The phase range is stored in non-volatile memory, and does not change when power has been off, after
a Factory Reset (*RST command), or after an Instrument Preset (SYSTem:PRESet command).
l The phase range is set to AUTO when the instrument is shipped from the factory or after a
SYSTem:SECurity:IMMediate command.
Return Format
The query returns the range: "AUTO", "POS" or "CENT".
Examples
The following example enables the 0° to 36 range for phase measurements.
FORM:PHAS POS
See Also
CONFigure:PHASe
MEASure:PHASe?
SENSe:FUNCtion
FORMat:PHASe
178 Keysight 53220A/53230A Programmer's Reference