EasyManua.ls Logo

Keysight 53230A - Sense:gate:stop:holdoff:source

Keysight 53230A
518 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SENSe:GATE:STOP:HOLDoff:SOURce
Syntax
[SENSe:]GATE:STOP:HOLDoff:SOURce {IMMediate | TIME | EVENts}
[SENSe:]GATE:STOP:HOLDoff:SOURce?
Description
This command selects whether time or input events (edges) are used to hold off enabling the gate close
(stop) source when advanced gating is enabled (SENSe:FREQuency:GATE:SOURce, SENSe:TINTer-
val:GATE:SOURce, or SENSe:TOTalize:GATE:SOURce set to ADVanced with the appropriate meas-
urement function selected).
Remarks
l Selecting IMMediate disables the holdoff.
l The holdoff starts when the gate opens for totalize measurements. For other measurements, the holdoff
starts on the first input event after the gate opens.
l If SENSe:GATE:STOP:SOURce is set to IMMediate, the gate will close immediately following the end of
the holdoff.
l If SENSe:GATE:STOP:SOURce is set to EXTernal, the gate will close on the next selected edge on the source
selected by SENSe:GATE:EXTernal:SOURce following the end of the holdoff. Edges that occur during the hol-
doff are ignored.
l The instrument sets the gate close holdoff source to IMMediate after a Factory Reset (*RST command) or an
Instrument Preset (SYSTem:PRESet command).
Return Format
The query returns the current gate close holdoff source: IMM, TIME or EVEN.
Examples
The following example makes a frequency measurement on channel 1. The gate opens .1 sec after a neg-
ative edge on the Gate In/Out BNC, and, after .25 seconds, closes on the next positive edge on the BNC.
CONF:FREQ 1.2E6
FREQ:GATE:SOUR ADV
GATE:STAR:SOUR EXT
GATE:STAR:SLOP NEG
GATE:STAR:DEL:SOUR TIME
GATE:STAR:DEL:TIME .1
GATE:STOP:SOUR EXT
SENSe:GATE:STOP:HOLDoff:SOURce
354 Keysight 53220A/53230A Programmer's Reference

Table of Contents