CALibration:LEVel3?
Syntax
CALibration:LEVel3? <frequency>, <power>
Description
This query performs a level calibration of channel 3 at the frequency specified. Before you can calibrate
the instrument, you must unsecure it by entering the correct security code.
For a more detailed discussion of the calibration procedures, see the Keysight
53210A/53220A/53230A 350 MHz Counter Service Guide. Please refer to the Ser-
vice Guide before attempting to calibrate the instrument. Improper use of the
CALibration commands can adversely affect the accuracy and reliability of the instru-
ment.
Parameters
Name Type Range of Values Default Value
<frequency> Numeric See the Keysight 53210A/53220A/53230A 350 MHz
Counter Service Guide
This is a required parameter
<power> Numeric See the Keysight 53210A/53220A/53230A 350 MHz
Counter Service Guide
This is a required parameter
Remarks
l
Refer to the Keysight 53210A/53220A/53230A 350 MHz Counter Service Guide for detailed pro-
cedures, including how to connect a calibration source, recommended equipment, the specified cal-
ibration points, and so forth.
l If a calibration fails, "+1" is returned and an error is stored in the error queue. For a complete listing of
the error messages related to calibration failures, see SCPI Error Messages.
l This query increments the calibration count on the counter (see CALibration:COUNt? command).
Return Format
The query returns +0 (level calibration passed) or +1 (level calibration failed).
Examples
The following example performs a level calibration and returns a pass/fail indication.
CAL:LEV3? 1.0E9,-27
Typical Response: +0
Keysight 53220A/53230A Programmer's Reference 103
CALibration:LEVel3?