EasyManua.ls Logo

Keysight 53230A - Page 127

Keysight 53230A
518 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remarks
l <expected> is the expected value of the ratio, not of either individual input frequency.
l
As defined in the SCPI standard, <resolution> must be specified in absolute frequency (hertz), not as a
fraction of the input signal frequency.
l The frequency of the first channel in the channel pair is used as the numerator of the ratio; the second
channel as the denominator.
l The input channels are configured for auto-leveling at 50% with a positive slope.
l The gate source (see SENSe:FREQuency:GATE:SOURCE) is set to TIME.
l
The counter uses the <expected> and <resolution> parameters to configure the setting for the
SENSe:FREQuency:GATE:TIME command. The table below shows the relationship between relative res-
olution and gate time in seconds, where relative resolution = <resolution> / <expected>.
Relative Resolution
Gate Time
(53220A)
Gate Time
(53230A)
1.1E-14 1000 sec 1000 sec
>1.1E-14 to 1.1E-13 1000 sec 100 sec
>1.1E-13 to 1.1E-12 100 sec 10 sec
>1.1E-12 to 1.1E-11 10 sec 1 sec
>1.1E-11 to 1.1E-10 1 sec 100 msec
>1.1E-10 to 1.1E-9 100 msec 10 msec
>1.1E-9 to 1.1E-8 10 msec 1 msec
>1.1E-8 to 1.1E-7 1 msec 100 usec
>1.1E-7 to 1.1E-6 100 usec 10 usec
>1.1E-6 100 usec 1 usec
l Use READ? or INITiate[:IMMediate] to start the measurement.
Examples
The following example configures the instrument for frequency ratio measurements between channel 1
and channel 2 using default parameters, makes two measurements, and then transfers the measurements
to the instrument output buffer.
CONF:FREQ:RAT
SAMP:COUN 2
READ?
Typical Response: +1.00230000000000E+001,+1.00190000000000E+001
The following example configures the instrument for 10-digit frequency ratio measurements between
channel 2 and channel 1, makes one measurement using an external trigger, and then transfers the meas-
urement to the instrument output buffer.
CONF:FREQ:RAT 1.2,1e-10,(@2),(@1)
TRIG:SOUR EXT;SLOP POS
Keysight 53220A/53230A Programmer's Reference 127
CONFigure:FREQuency:RATio

Table of Contents

Other manuals for Keysight 53230A

Related product manuals