Device Clear performs the following actions:
l If a measurement is in progress, it is aborted.
l The instrument returns to the trigger "idle" state.
l The instrument's input and output buffers are cleared.
l The instrument is prepared to accept a new command string.
l An overlapped command, if any, will be terminated with no "Operation Complete" indication (applies to the
INIT command).
The ABORt command is the recommended method to terminate a measurement.
Introduction to the SCPI Language
16 Keysight 53220A/53230A Programmer's Reference