*CAL?
Syntax
*CAL?
Description
This query performs an auto-calibration on the instrument and returns a result indicating whether the
auto-calibration succeeded (+0) or failed (+1).
Remarks
l If the auto-calibration fails, use SYSTem:ERRor? get more details concerning the failure.
l This command takes about 10 seconds to execute.
Return Format
The query returns +0 if calibration passes, +1 if it fails.
Examples
The following example performs an auto-calibration:
*CAL?
Typical Response: +0
See Also
CALibration Subsystem
SYSTem:TEMPerature?