*ESE
Syntax
*ESE <enable>
*ESE?
Description
This command enables bits in the enable register for the Standard Event Register group. The selected
bits are then reported to bit 5 of the Status Byte Register.
For more information on the SCPI Status System, see Status System Introduction.
Parameters
Name Type Range of Values Default Value
<enable> Numeric A decimal value which corresponds to the
binary-weighted sum of the bits in the
register (see table below).
This is a required parameter
Remarks
l The following table lists the bit definitions for the Standard Event Register.
Bit
Number
Bit Name Decimal
Value
Definition
0
Operation Com-
plete
1 All commands prior to and including *OPC have
been executed.
1 Not Used 2 (Reserved for future use)
2 Query Error 4 The instrument tried to read the output buffer
but it was empty. Or, a new command line was
received before a previous query has been read.
Or, both the input and output buffers are full.
3 Device-Specific
Error
8 A device-specific error, including a self-test error
or calibration error, occurred (an error in the -300
range or any positive error has been generated).
For a complete listing of the error messages, see
SCPI Error Messages.
4 Execution Error 16 An execution error occurred (an error in the -200
range has been generated).
5 Command Error 32 A command syntax error occurred (an error in the
-100 range has been generated).