Loading...
*OPC
Syntax
*OPC
Description
This command sets the "Operation Complete" bit (bit 0) in the Standard Event register at the completion
of the current operation.
For more information on the SCPI Status System, see Status System Introduction.
Remarks
l The purpose of this command is to synchronize your application with the instrument.
l Note the difference between the *OPC command and the *OPC? query command. The latter returns 1
to the output buffer at the completion of the current operation.
Examples
The following example sets the "Operation Complete" bit.
*OPC
See Also
*OPC?
*OPC
188 Keysight 53220A/53230A Programmer's Reference