*RCL
Syntax
*RCL {0 | 1 | 2 | 3 | 4}
Description
This command recalls the instrument state stored in the specified storage location. You cannot recall the
instrument state from a storage location that is empty or was deleted. When shipped from the factory,
storage locations 1 through 4 are empty (location 0 has the power-on state).
Use the *SAV command to store the current instrument state.
Remarks
l The instrument has five storage locations in non-volatile memory to store instrument states. The instru-
ment uses location 0 to automatically hold the state of the instrument at power off. You can only recall
a state from a location that contains a previously stored state.
l The instrument states are stored in non-volatile memory and do not change when power has been off,
after a Factory Reset (*RST command), or after an Instrument Preset (SYSTem:PRESet command).
l The MEMory:STATe:DELete and MEMory:STATe:DELete:ALL commands delete, but do not sanitize,
stored states.
l SYSTem:SECurity:IMMediate command deletes and sanitizes all stored states.
Examples
The following example recalls the instrument state previously stored in location 1.
*RCL 1
See Also
*SAV
MEMory:STATe:DELete
MMEMory:LOAD:STATe
MMEMory:STORe:STATe
*RCL