>1.1E-9 to 1.1E-8 10 msec 1 msec
>1.1E-8 to 1.1E-7 1 msec 100 usec
>1.1E-7 to 1.1E-6 100 usec 10 usec
>1.1E-6 100 usec 1 usec
Return Format
The query returns a single measurement. The format of the measurement is controlled by the FORMat
Subsystem commands.
Examples
The following example configures the instrument for pulse repetition frequency measurements on chan-
nel 1 using default parameters, makes one measurement, and then transfers the measurement to the
instrument output buffer.
MEAS:FREQ:PRF?
Typical Response: +1.00230000000000E+007
The following example configures the instrument for 10-digit pulse repetition frequency measurements of
an approximately 1.2 MHz pulsed signal on channel 2, makes one measurement, and then transfers the
measurement to the instrument output buffer.
MEAS:FREQ:PRF? 1.2E6,1e-4,(@2)
Typical Response: +1.21513398300000E+006
See Also
CONFigure?
FETCh?
READ?
SENSe:FREQuency:GATE:SOURce
SENSe:FREQuency:GATE:TIME
MEASure:FREQuency:PRF?
248 Keysight 53220A/53230A Programmer's Reference