EasyManua.ls Logo

Keysight 53230A - Page 255

Keysight 53230A
518 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Return Format
The query returns a single measurement. The format of the measurement is controlled by the FORMat
Subsystem commands.
Examples
The following example configures the instrument for fall time measurements on channel 1 using default
parameters, makes one measurement, and then transfers the measurement to the instrument output buf-
fer.
MEAS:FTIM?
Typical Response: +5.10520000000000E-006
The following example configures the instrument for fall time measurements on channel 2 with 20% and
80% reference levels, makes one measurement, and then transfers the measurement to the instrument
output buffer.
MEAS:FTIM? 20,80,(@2)
Typical Response: +5.02573000000000E-006
See Also
CONFigure?
FETCh?
READ?
Keysight 53220A/53230A Programmer's Reference 255
MEASure:FTIMe?

Table of Contents

Other manuals for Keysight 53230A

Related product manuals