Return Format
The query returns a single measurement. The format of the measurement is controlled by the FORMat
Subsystem commands.
Examples
The following example configures the instrument for rise time measurements on channel 1 using default
parameters, makes one measurement, and then transfers the measurement to the instrument output buf-
fer.
MEAS:RTIM?
Typical Response: +5.10520000000000E-006
The following example configures the instrument for rise time measurements on channel 2 with 20% and
80% reference levels, makes one measurement, and then transfers the measurement to the instrument
output buffer.
MEAS:RTIM? 20,80,(@2)
Typical Response: +5.02573000000000E-006
See Also
CONFigure?
FETCh?
READ?
MEASure:RTIMe?
272 Keysight 53220A/53230A Programmer's Reference