Return Format
The query returns the frequency gate source: TIME, EXT, INP, INP2 or ADV.
Examples
The following example makes an internally-gated frequency measurement on channel 1 with a 1 second
gate time:
CONF:FREQ 1.2E6
FREQ:GATE:SOUR TIME
FREQ:GATE:TIME 1.0
READ?
Typical Response: +1.21513398300000E+006
See Also
SENSe:FREQuency:GATE:TIME
SENSe:FREQuency:GATE:POLarity
SENSe:GATE:STARt:SOURce
SENSe:GATE:STARt:SLOPe
SENSe:GATE:STARt:DELay:SOURce
SENSe:GATE:STARt:DELay:EVENts
SENSe:GATE:STARt:DELay:TIME
SENSe:GATE:STOP:SOURce
SENSe:GATE:STOP:SLOPe
SENSe:GATE:STOP:HOLDoff:SOURce
SENSe:GATE:STOP:HOLDoff:EVENts
SENSe:GATE:STOP:HOLDoff:TIME
Keysight 53220A/53230A Programmer's Reference 325
SENSe:FREQuency:GATE:SOURce