10 Instrument Locked 1024 If a remote interface (GPIB, USB, or LAN) has a lock (see
SYSTem:LOCK:REQuest? command), this bit will be set.
When a remote interface releases the lock (see
SYSTem:LOCK:RELease command), this bit will be cleared.
11 Configuration Change 2048 Instrument configuration has been changed, either from the front
panel or from SCPI.
12 Memory Threshold 4096 Programmed number of measurements have been stored in reading
memory (see DATA:POINts:EVENt:THReashold command).
13 Global Error 8192 Set if any remote interface has an error in its error queue; cleared oth-
erwise.
14 Not Used 16384 (Reserved for future use)
15 Not Used 32768 "0" is returned.
l To enable specific bits, you must write a decimal value that corresponds to the binary-weighted sum of
the bits in the register. The selected bits are summarized in the "Standard Operation Summary" bit (bit
7) of the Status Byte Register.
l Use the <enable_value> parameter to specify which bits will be summarized to the Status Byte Register.
The decimal value specified corresponds to the binary-weighted sum of the bits you wish to enable in
the register. For example, to enable bit 5 (decimal value = 32) and bit 9 (decimal value = 512), the cor-
responding decimal value would be 544 (32 + 512).
l The *CLS (clear status) command will not clear the enable register but it does clear all bits in the event
register.
l The STATus:PRESet command will clear all bits in the enable register.
l The enable register is unaffected by a Factory Reset (*RST command), or an Instrument
Preset (SYSTem:PRESet command). It is set to +32767 at power-on.
Return Format
The query reads the enable register and returns a decimal value which corresponds to the binary-
weighted sum of all bits set in the register. For example, if bit 4 (decimal value = 16) and bit 9 (decimal
value = 512) are enabled, the query will return +528.
Examples
The following example enables bit 5 (decimal value = 32) in the enable register.
STAT:OPER:ENAB 32
The following example returns which bits are enabled in the register.
STAT:OPER:ENAB?
Typical Response: +32