Loading...
l For the query form, specify "CURRent" (default) to read the dynamic default gateway currently being
used by the instrument. Specify "STATic" to read the default gateway currently stored in non-volatile
memory within the instrument (may not be the actual address used by the instrument if DHCP is enabled).
l The default gateway is stored in non-volatile memory, and does not change when power has been off,
after a Factory Reset (*RST command), or after an Instrument Preset (SYSTem:PRESet command).
l The gateway is set to "0.0.0.0" (no subnetting) when the instrument is shipped from the factory or after a
SYSTem:SECurity:IMMediate command.
Return Format
The query returns the default gateway address in the form "nnn.nnn.nnn.nnn" (the quotes are also
returned).
Examples
The following example sets the default gateway address.
SYST:COMM:LAN:GAT "198.105.232.1"
SYST:COMM:LAN:UPD
The following example returns the default gateway address currently being used by the instrument.
SYST:COMM:LAN:GAT? CURR or SYST:COMM:LAN:GAT?
Typical Response: "0.0.0.0"
See Also
SYSTem:COMMunicate:LAN:DHCP
SYSTem:COMMunicate:LAN:IPADdress
SYSTem:COMMunicate:LAN:SMASk
Keysight 53220A/53230A Programmer's Reference 411
SYSTem:COMMunicate:LAN:GATeway