ASAP 2460 Operator’s Manual Analysis Conditions Files
246-42800-01 - Aug 2013 3-17
Isotherm Collection
group box
Specify how isotherms are to be collected: Target Pressures or Dose
Increments.
Target Pressures
option
Select to have isotherms collected by specifying target pressures. Select this
option to enable the Pressures and Options buttons.
Pressures button - click to edit the Entered Pressures table. Use this
window to manually enter the pressures.
A pressure table consists of relative pressure points at which data
(isotherms) are to be collected. The relative pressures may span the entire
range of 0.00000001 to 0.995 p/p
o
. There must be one adsorption branch
followed optionally by one desorption branch.