Appendix B ASAP 2460 Operator’s Manual
B-20 246-42800-01 - Dec 2013
4022- The entered bath temperature value (Po and Temperature Options of the Anal-
ysis Conditions) is outside the range of the temperatures listed in the Psat vs.
T
emperature Table (Adsorptive Properties).
Cause: The Po that corresponds to the entered temperature cannot be found
because the entered temperature is outside of the range of values in the
Psat vs. T table.
Action A: Enter a temperature that is within
the range of the table.
Action B: Use a different Psat vs. T table.
4023- The file (sample file) cannot be prepared for analysis. It is open for editing and
contains errors.
Cause: You attempted an analysis using a sample file that is being edited and
has invalid values in one or more of its fields.
Action: Enter valid values into the fields that have e
rrors, save the changes, and
close the dialog box.
4024- Backfill gas in sample file does not match any gas in unit.
Cause: You selected a backfill gas that is not included in the gas configuration.
Action: Make sure that gas selections match the gases that are attached to the
inst
rument. Choose one of the gases as the backfill gas, or edit the sam-
ple file and choose one of the existing
gases as the backfill gas.
4025- There is no Helium attached to the unit.
Cause: A measured free space has been requested but helium does not appear
to be available.
Action: Make sure the helium cylinder is attached to the
unit at its designated
port.