Appendix B ASAP 2460 Operator’s Manual
B-22 246-42800-01 - Dec 2013
4031- Not enough points with a relative pressure in the range (n,n)
Cause: Fewer than two data points selected for the Dubinin report falls within
the selected relative pressure range.
Action: Edit the calculation assignments or the fitted
relative pressure range.
4033- Not enough points to generate Dubinin Tabular Report.
Cause: There are fewer than two valid data points available for the Dubinin
tabular reports.
Action: Examine the calculation assignments in the
Collected Data dialog of
the sample file. You must have at least two micropore pressures
selected for inclusion in the Dubinin report.
4034- Fewer than 2 points available for Dubinin calculations.
Cause: There are fewer than two valid data points available for Dubinin
reports in one of the sample files selected for overlaying.
Action: Examine the calculation assignments in the
Collected Data dialog of
the sample file. You must have at least two micropore pressures
selected for inclusion in the Dubinin report.
4035- Cannot calculate optimized Astakhov exponent: Not enough points with a rel-
ative pressure in the range [(pressure), (pressure)].
Cause: There are fewer than two valid data points in the relative pressure
range specified. Astakhov reports will not be produced.
Action: Examine the calculation assignments in the Collec
ted Data dialog for
the sample file. You must have at least two pressures in the given range
selected for inclusion in the Dubinin report.