Appendix B ASAP 2460 Operator’s Manual
B-26 246-42800-01 - Dec 2013
4045- Points in the report pressure table lie outside the collected data.
Cause: Calculation assignments are not being used and more than one of the
report pressure table points is above the range of the collected data, and
more than one is below.
Action: Change the report pressure table to
be more consistent with the col-
lected data.
4046- (file name) could not be opened for reading.
Cause: A thickness curve file could not be opened.
Action: If the problem persists, restart your computer and optionally perform a
me
dia integrity check (using ScanDisk).
4047- Warning: An error occurred while reading (file name).
Cause: An error happened during a read operation of a thickness curve file.
Action: If the problem persists, restart your computer and optionally perform a
me
dia integrity check (using ScanDisk).
4050- None of the requested quantities adsorbed is within the range of the collected
data of more than one sample file.
Cause: The files selected for HOA reporting cannot produce a valid report.
Action: Choose a different set of files.
4051- The sample (file name) does not have any data in the range of the requested
quantities adsorbed.
Cause: The sample’s data cannot be interpolated to any of the quantities
adsorbed.
Action: Edit the quantities adsorbed so that they are
within the range of the col-
lected data.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics ASAP 2460 and is the answer not in the manual?

Micromeritics ASAP 2460 Specifications

General IconGeneral
BrandMicromeritics
ModelASAP 2460
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals