ASAP 2460 Operator’s Manual Appendix B
246-42800-01 - Dec 2013 B-39
6247- Warm Freespace failure: P1 = (pressure), P2 = (pressure
Cause A: Helium gas line is not properly connected to the analyzer.
Action A: Check the helium gas connections.
Cause B: A valve, transducer, or elevator has failed.
Action B: Contact your Micromeritics representative.
).
6248- Cold Freespace failure: P3 = (pressure).
Cause A: Helium gas line is not properly connected to the analyzer.
Action A: Check the helium gas connections.
Cause B: A valve, transducer, or elevator has failed.
Action B: Contact your Micromeritics representative.
6249- Zeroing of a transducer failed. Analysis Canceled.
Action A: Set the Psat gas regulator to 15 psig (1.03 bar). Then restart the
analysis.
Cause B: The Psat gas bottle is empty.
Action B: Connect a new Psat gas bottle. Then restart the analysis.
Caus9e C: The Psat tube fitting is loose.
Action C: Check the Psat tube fittings and ensure
that the tube is attached
securely to the port. Then restart the analysis.
Cause: The transducer did not respond correctly.