EasyManuals Logo

Micromeritics ASAP 2460 User Manual

Default Icon
468 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #357 background imageLoading...
Page #357 background image
ASAP 2460 Operator’s Manual Appendix C
246-42800-01 - Aug 2013 C-33
The following values were used to calculate the values in Table C-1.
D
A
values are from van der Waal’s constant.
N
A
values are from liquid densities.
and values are derived from data found in Ross and Olivier
12
.
The physical parameters referenced in Saito/Foley are:
Carbon-Graphite Zeolite
D
S
=3.40 D
S
=3.04
N
S
= 3.845 × 10
15
N
S
=3.75 × 10
15
S
= 1.05 × 10
-29
(Ross/Olivier)
S
= 1.94 × 10
-29
13.5 × 10
-29
(Horvath/Kawazoe, implicit)
S
=0.85 × 10
-24
S
=1.02 × 10
-24
Nitrogen Argon
D
A
=3.00 D
A
=2.95
A
=6.71 × 10
14
A
= 7.608 × 10
14
A
= 3.6 × 10
-29
A
=3.22 × 10
-29
A
=1.76 × 10
-24
A
=1.63 × 10
-24
Carbon Dioxide
D
A
=3.23
N
A
= 4.567 × 10
14
(25 ºC)
5.45 × 10
14
(0 ºC)
7.697 × 10
14
(-78 ºC)
A
=5.0 × 10
-29
A
=2.7 × 10
-24
Aluminophosphate Aluminosilicate
D
S
= 2.60 D
S
=2.76
N
S
=1.48 × 10
15
N
s
=1.31 × 10
15
S
= 1.3 × 10
-29
S
=1.3 × 10
-29
S
=2.5 × 10
-24
S
=2.5 × 10
-24

Table of Contents