4422 The file %s does not exist.
Cause: The selected file does not exist on the disk drive.
Action: Select an existing file. Ensure that the file has been created before use.
4423 The sample %s is already selected on port %d.
Cause: The selected sample file is already selected for use on a different port.
Action: Select another sample file for this port.
4424 The file %s on port %d could not be opened. Check if the sample file is
already in use for editing or analysis.
Cause: The selected sample file is already open by this or another application.
Cause: The selected sample file is damaged.
Action: Select another sample file.
4425 The sample %s on port %d has an invalid status and cannot be used
for degassing.
Cause: The status of the file is not consistent with the current operation.
Action: Select a sample file that has not been used for an analysis. Only sample files with a
status of No Analysis or Prepared may be selected.
4426 Port %d is currently in use. Operation cannot be started.
Cause: The current operation cannot be completed because the port is already in use.
Action: Wait for port to terminate operation or perform the desired operation on an unused port.
8 Error Messages
Smart VacPrep Operator Manual 8 - 3

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics ASAP 2460 and is the answer not in the manual?

Micromeritics ASAP 2460 Specifications

General IconGeneral
BrandMicromeritics
ModelASAP 2460
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals