EasyManua.ls Logo

Mobius AIRO Mobile CT System - Page 190

Mobius AIRO Mobile CT System
294 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
178 MI-42-0001 Rev. 19
QUALITY TESTING: Testing Image Quality
NOTES:
AIRO will automatically scan the length of the Phantom, process the data, and run the Quality
Checks.
QC Image Data is automatically stored on AIRO under Existing Scans with a patient name of
Phantom.
Options
Retry after a Length of Travel warning:
If the QC scan terminates and the System Warning dialog indicates that scan was longer than
the available length of travel:
1. Touch
OK
.
2. On the Pendant, move through the work flow for the terminated scan by touching
Next
on:
QC Failed result summary screen
3D-image viewer screen
Select DICOM node screen
NOTE: Because the QC scan terminated, there is no significant data on these screens.
3. On the dialog asking to continue without sending results to a DICOM node, select
OK
.
See Figure 146.
4. Move the Gammex 464 Phantom closer to the center of the base along the z-axis.
5. Re-align it with the laser lines (horizontal and vertical) as shown in Figure 140.
6. Return to the start of the QC cycle:
For systems with the Pediatrics feature enabled, go to the Select KV screen and
restart the QC scan cycle. See Figure 139.
For systems with the Pediatrics feature disabled, go to the Position the Gammex
screen and restart the QC scan cycle. See Figure 140.
Retry after an Endpoint Set Failure:
If a QC scan fails to start and the Endpoint Set Failure dialog appears:
1. Touch
OK
to return to the Position Gammex screen.
2. Move the Gammex 464 Phantom closer to the center of the base along the z-axis.
3. Re-align it with the laser lines (horizontal and vertical) as shown in Figure 140.
4. Touch
OK
to run the QC scan.

Table of Contents