Instrument Function
R&S
®
SMBV100A
131Operating Manual 1407.6062.32 ─ 08
If this parameter is disabled, the instrument cannot be operated manually via the user
interface, i.e. display, front panel keys and external keyboard are disabled. Remote con-
trol of the insrument is enabled.
The setting requires the entry of the security password and is only accepted after the
"Accept" button is pressed.
SCPI command:
:SYSTem:DLOCk on page 712
Keyboard
Enables/disables an external keyboard and mouse connected to the instrument, the front
pannel keys of the instrument, the rotary knob and the on-screen keyboard.
If this parameter is disabled, the instrument cannot be manually controlled but changes
in the settings are shown on the display. Remote access and remote control are enabled.
The setting requires the entry of the security password and is only accepted after the
"Accept" button is pressed.
To enable the keyboard, type the security password on the external keyboard or use the
SYST:KLOC OFF command over remote control.
SCPI command:
:SYSTem:KLOCk on page 715
Security Password
Enters the password that is required to enable or to disable the settings protected by a
security password. Default is '123456'.
Note: It is highly recommended to change the default security password before connect-
ing the instrument to the network.
All settings are only accepted after the "Accept" button is pressed.
SCPI command:
n.a.
Accept - Security
Accepts a new entry or selection and changes the settings accordingly.
SCPI command:
n.a.
Bluetooth Pin
Requires a USB Bluetooth adapter (recommended extra, see data sheet).
Enters the Bluetooth pin of an external Bluetooth device. The pin is required to enable
remote control via an external Bluetooth device.
The interface parameters are identical to the RS232 interface parameters and can be set
in the "Remote Channel Settings" dialog (see "RS232 using USB
adapter" on page 126).
SCPI command:
n.a.