Remote Control Commands
R&S
®
SMBV100A
554Operating Manual 1407.6062.32 ─ 08
Parameters:
<Mode> NONE | MASTer | SLAVe
NONE
The instrument is working in stand-alone mode.
MASTer
The instrument provides all connected instrument with its syn-
chronisation (including the trigger signal) and reference clock sig-
nal.
SLAVe
The instrument receives the synchronisation and reference clock
signal from another instrument working in a master mode.
*RST: NONE
Example:
BB:DM:CLOC:SYNC:MODE MAST
the instrument is configured to work as a master one.
[:SOURce<hw>]:BB:DM:CLOCk:SYNChronization:EXEcute
Performs automatically adjustment of the instrument's settings required for the synchro-
nization mode, set with the command BB:DM:CLOC:SYNC:MODE.
Example:
BB:DM:CLOC:SYNC:MODE MAST
the instrument is configured to work as a master one.
BB:DM:CLOC:SYNC:EXEC
all synchronization's settings are adjusted accordingly.
Usage: Event
Trigger Settings
The numeric suffix to OUTPut distinguishes between the available markers.
[:SOURce<hw>]:BB:DM:TRIGger:ARM:EXECute..............................................................554
[:SOURce<hw>]:BB:DM:TRIGger:EXECute......................................................................555
[:SOURce<hw>]:BB:DM:TRIGger:EXTernal:SYNChronize:OUTPut.....................................555
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut:DELay:FIXed....................................................556
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:DELay......................................................556
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:DELay:MINimum.......................................557
[:SOURce<hw>]:BB:DM:TRIGger:OUTPut<ch>:DELay:MAXimum......................................557
[:SOURce<hw>]:BB:DM:TRIGger:RMODe........................................................................557
[:SOURce<hw>]:BB:DM:TRIGger:SLENgth.......................................................................558
[:SOURce<hw>]:BB:DM:TRIGger:SOURce.......................................................................558
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal<ch>]:DELay..................................................559
[:SOURce<hw>]:BB:DM:TRIGger[:EXTernal<ch>]:INHibit..................................................559
[:SOURce<hw>]:BB:DM[:TRIGger]:SEQuence..................................................................559
[:SOURce<hw>]:BB:DM:TRIGger:ARM:EXECute
The command stops signal generation for trigger modes Armed_Auto and Armed_Ret-
rigger. A subsequent internal or external trigger event restart signal generation.
SOURce Subsystem