26 Number of PDOs
The Number of PDOs parameter is used to determine the number of process data
objects (PDOs) that transmit the reading result.
When reading results are transmitted via PDOs, the reading result data is distributed
across several PDOs. The PDOs are transmitted in sequence. The configured reading
result is restricted to 50 characters for this type of transmission.
27 Number of Attempts
Enter the maximum number of read attempts that are to be made with an activated
parameter set in the event of an automatic parameter switchover under the Number of
Attempts parameter.
28 Application Identifier Marking
If the Application Identifier Marking parameter is activated, the APPLICATION IDENTIFIERS
appear in parentheses in the output format.
On the basis of the APPLICATION IDENTIFIERS, the subsequent values of the code content are
assigned to a defined property. This means that the data contained within a code (e.g.,
use-by date, batch number, date of manufacture, etc.) can be represented in a standar‐
dized format.
29 Application Counters
Events such as successful (GOOD READ) and unsuccessful (NO READ) read operations can
be counted and saved on the Application Counters tab.
Depending on the setting, the device stores the data temporarily in the RAM or perma‐
nently in the flash memory.
Quantitative conditions are defined using counters and can be used for output format‐
ting, for example. Counting functions can be used to control processes, e.g., object bun‐
dling is initiated when a defined counter value is reached.
30 APS/EPS 1
Camera and illumination settings that are used cyclically in the event of automatic or
event-controlled parameter switchover can be made in the APS/EPS 1 to APS/EPS 4
groups.
Pressing the APPLY STANDARD PARAMETERS button applies the parameter settings saved
under the CAMERA AND ILLUMINATION tab.
31 Output Condition
If the Data transmission point parameter is set to the value AS SOON AS POSSIBLE, the Out‐
put Condition parameter can be used to select the data output condition.
•
GOOD READ: The reading results are output if the read operation was successful.
•
MATCH1 CONDITION: The reading results are output if the code that has been read cor‐
responds to the MATCH1 CONDITION match code.
•
MULTICODES1 CONDITION: The reading results are output if the number of codes that
have been read corresponds to the number specified by the MULTICODES1 CONDITION.
•
TEACH-IN1 CONDITION: The reading results are output if the code that has been read
corresponds to the TEACH-IN1 CONDITION match code.
•
TEACH-IN2 CONDITION: The reading results are output if the code that has been read
corresponds to the TEACH-IN2 CONDITION match code.