210 Inactive
The Inactive parameter is used to select the event that leads to the digital output being
deactivated.
TIMER/TRACKING: The output is deactivated on expiry of a specific period (Length).
NEXT OBJECT TRIGGER: The output is deactivated as soon as the next reading gate
opens.
SOPAS COMMAND: The output can be deactivated via a command. This allows defined
statuses to be displayed by a connected indicator lamp when a corresponding
command is issued by the control (PLC).
GOOD READ: The output is deactivated if the read operation was successful.
NO READ: The output is deactivated if the read operation was unsuccessful.
TEACH-IN 1START: The output is deactivated as soon as the MATCH CODE TEACH-IN 1 teach-
in process commences.
TEACH-IN 2START: The output is deactivated as soon as the MATCH CODE TEACH-IN 2 teach-
in process commences.
MATCH 1 CONDITION: The output is deactivated if the code that has been read corre‐
sponds to the MATCH 1 CONDITION match code.
MULTICODES1 CONDITION: The output is deactivated if the number of codes that have
been read corresponds to the number specified by the MULTICODES1 CONDITION.
TEACH-IN1 CONDITION: The output is deactivated if the code that has been read corre‐
sponds to the TEACH-IN1 CONDITION match code.
TEACH-IN2 CONDITION: The output is deactivated if the code that has been read corre‐
sponds to the TEACH-IN2 CONDITION match code.
211 Inhibit Time
The Inhibit Time parameter defines the interval that must elapse between the repeated
transmission of process data objects (PDOs) in ms.
212 Increment
The Increment group contains parameters for converting distance and speed signals.
The signals are either generated by the configuration software (SOPAS)/CAN interface
or via an incremental encoder at the digital input. Incremental encoders are sensors for
detecting positional changes (linear) or angle changes (rotating) that can capture the
distance and direction or the angle change and direction of rotation. They are also
known as rotary encoders, incremental rotary encoders, and rotary pulse encoders.
213 Increment Configuration
The Increment Configuration tab contains parameters for converting distance and
speed signals.
The signals are either generated by the configuration software (SOPAS)/CAN interface
or via an incremental encoder at the digital input. Incremental encoders are sensors for
detecting positional changes (linear) or angle changes (rotating) that can capture the
distance and direction or the angle change and direction of rotation. They are also
known as rotary encoders, incremental rotary encoders, and rotary pulse encoders.
SOPAS PARAMETER OVERVIEW
78
O N L I N E H E L P | Lector62x 8013778/YPV6/2016-01-14 | SICK
Subject to change without notice

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SICK Lector62 Series and is the answer not in the manual?

SICK Lector62 Series Specifications

General IconGeneral
BrandSICK
ModelLector62 Series
CategoryBarcode Reader
LanguageEnglish

Related product manuals