SVAN 979 User Manual - Appendixes
Group name
#1
code
#7
code
#S
code
Code value description
mt
#7,mt,YYYY-MM-
DD,HH:MM:SS,Tx.x,Px,Hx.x,Vx.x,Dx,Rx,Nx.x;
where:
YYYY-MM-DD results date
HH:MM:SS - results time
Tx.x - temperature in °C
Px - pressure in hPa
Hx.x - humidity in %
Vx.x - wind speed in m/s
Dx - wind direction in °
Rx - rain flag (0 or 1)
Nx.x - rain intensity mm/h
Send screen (read only)
SP
#7,SP,x; where x:
1 - function returns the screen settings
2 - function returns the color palette
3 - function returns the screen status
External cable connection
(read only)
PC
0 - disconnected
1 - connected
Alarm (read only)
AL
#7,AL; This function returns the alarm status in
the format #7,AL,curr_level,text,alarm_level;
where:
curr_level result that triggered the alarm
text text of the alarm
alarm_level the value of the trigger