SVAN 979 User Manual - Appendixes
triggering source (parameter relevant only to modes in which
triggering is related to signal level):
0 – triggering by RMS/Leq level from the first profile,
– remaining values reserved
triggering threshold value in 0.01 dB format (parameter relevant only
for modes in which triggering is related to signal level)
recording period before triggering condition occurs in the 10 ms
format (parameter relevant only for recording on trigger and in alarm
mode)
triggering status check period in 0.1 ms format - zero means period
equal to logger step (parameter relevant only for recording with
triggering)
signal rate Sample in 10 Hz format
maximum recording time in seconds - zero means no registration
time limit (parameter relevant only for recording with triggering)
measurement filter for the recorded signal:
0 – HP
1 – Z
2 – A
3 – C
5 – B
number of bits per sample
marker triggering (parameter relevant only to triggering mode):
0 – disabled
1 – enabled
Table B.1.12. Parameters of external port settings (Multifunction I/O)
2Eh = block identifier
0nnh = block length in words
port mode:
0 – analog output (Analog Out)
1 – digital input (Digital In),
2 – digital output (Digital Out)
port function for digital input mode:
0 – external triggering input
port function for digital output mode:
0 – external triggering pulse output
1 – alarm pulse output
impulse active level (parameter relevant only to alarm pulse function
in digital output mode):
0 – low active level (Low)
1 – high active level (High)