EasyManua.ls Logo

TELEDYNE API T703U - Photometer Span Calibration

Default Icon
273 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration and Verification Teledyne API T703/T703U Calibrator Operation Manual
124
Photometer Span Calibration 6.1.3.2.
To set the response SLOPE for the T703 Photometric Calibrator‟s photometer,
press:
SETUP X.X O3 GAS CONFIG
BCAL DARK EXIT
Allow calibration gas to gas to enter the
PHOTOMETER IN port at the rear of the
calibrator.
Wait
A MINIMUM
OF
10 MINUTES
SETUP X.X O3 GAS CONFIG
MODE ADJ PHOT EXIT
SYSTEM ENTER PASSWORD:0
0 0 0 ENTR EXIT
Toggle these buttons until
PASSWORD = 717
SETUP X.X O3 PHOTOMETER BENCH CAL
EDIT CAL EXIT
SETUP X.X BENCH CAL:XSPN
ZERO SPAN XZRO XSPN ENTR EXIT
BCAL XSPN [TEST FUNCTION Message Field]
<TST TST> ZERO SPAN EXIT
The Message field displays
the Test Function currently
selected
Press YES to changes the
OFFSET & SLOPE values
for the O
3
measurement.
Press NO to leave the
photometer calibration
values unchanged.
BCAL XSPN [TEST FUNCTION Message Field]
<TST TST> ZERO SPAN EXIT
BCAL XSPN ARE YOU SURE?
YES NO
It is recommended
that you use an O
3
Span concentration
of 400 PPB
BCAL XSPN ACTUAL CONC: 400.0 PPB
0 0 4 0 0 .0 ENTR EXIT
Toggle these buttons to enter
the exact span cal value as
measured by the reference
photometer
(see figures 8-3 and 8-4)
EXIT discards
the new setting
ENTR accepts
the new setting
STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ SETUP
SETUP X.X PRIMARY SETUP MENU
O3 SEQ CFG CLK PASS MORE EXIT
Press SPAN if using the
Internal
O
3
Generator
Press XSPN if using an
External
O
3
Generator
07223D DCN7107

Table of Contents