EasyManua.ls Logo

TELEDYNE API T703U - Manual Calibration of the TEST Channel Configured for Voltage Ranges; Figure 4-3: Setup for Calibrating the TEST CHANNEL; Table 4-10: Voltage Tolerances for the TEST CHANNEL Calibration

Default Icon
273 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Teledyne API T703/T703U Calibrator Operation Manual Operating the Calibrator
79
Manual Calibration of the TEST Channel Configured for Voltage Ranges 4.9.2.3.
For highest accuracy, the voltages of the analog outputs can be manually
calibrated.
NOTE:
The menu for manually adjusting the analog output signal level will only appear if the AUTO-CAL
feature is turned off for the channel being adjusted (see Section 4.9.2.1)
Calibration is performed with a voltmeter connected across the output terminals
(See Figure 4-2) and by changing the actual output signal level using the front
panel menu buttons in 100, 10 or 1 count increments.
V
+DC Gnd
Recording
Device
V IN +
V IN -
ANALYZER
V OUT +
V OUT -
250 Ω
Volt
Meter
Figure 4-3: Setup for Calibrating the TEST CHANNEL
Table 4-10: Voltage Tolerances for the TEST CHANNEL Calibration
FULL
SCALE
ZERO
TOLERANCE
SPAN VOLTAGE
SPAN
TOLERANCE
MINIMUM
ADJUSTMENT
(1 count)
0.1 VDC
±0.0005V
90 mV
±0.001V
0.02 mV
1 VDC
±0.001V
900 mV
±0.001V
0.24 mV
5 VDC
±0.002V
4500 mV
±0.003V
1.22 mV
10 VDC
±0.004V
4500 mV
±0.006V
2.44 mV
07223D DCN7107

Table of Contents