Teledyne API T703/T703U Calibrator Operation Manual Principles of Operation
193
Figure 9-9: Location of the AC Configuration Jumper for the Dry Air Pump
AC Configuration – Internal Pump (JP7) 9.2.6.1.
AC power configuration for the internal dry air pump is set using Jumper set JP7.
Table 9-2: AC Power Configuration for Internal Pumps (JP7)
Connects pump pin 3 to 110 / 115 VAC power line
Connects pump pin 3 to 110 / 115 VAC power line
Connects pump pins 2 & 4 to Neutral
Connects pump pin 3 to 110 / 115 VAC power line
Connects pump pin 3 to 110 / 115 VAC power line
Connects pump pins 2 & 4 to Neutral
Connects pump pins 3 and 4 together
Connects pump pin 1 to 220 / 240VAC power line
Connects pump pins 3 and 4 together
Connects pump pin 1 to 220 / 240VAC power line
1
A jumper between pins 5 and 10 may be present on the jumper plug assembly, but is only functional on the T300 and
has no function on the Models T700 or T703.