19
3. GETTING STARTED
3.1. UNPACKING AND INITIAL SETUP
CAUTION
THE T703 WEIGHS ABOUT 16.1 KG (35.5 POUNDS) WITHOUT OPTIONS
INSTALLED. TO AVOID PERSONAL INJURY, WE RECOMMEND USING TWO
PERSONS TO LIFT AND CARRY THE CALIBRATOR.
NOTE
It is recommended that you store shipping containers/materials for future use if/when the
instrument should be returned to the factory for repair and/or calibration service. See Warranty
section in this manual and shipping procedures on our Website at http://www.teledyne-api.com
under Customer Support > Return Authorization.
1. Inspect the received packages for external shipping damage. If damaged,
please advise the shipper first, then Teledyne API.
2. Included with your calibrator is a printed record of the final performance
characterization performed on your instrument at the factory. This record,
titled Final Test and Validation Data Sheet (P/N 05760) is an important
quality assurance and calibration record for this instrument. It should be
placed in the quality records file for this instrument.
3. Carefully remove the top cover of the calibrator and check for internal
shipping damage.
Remove the set-screw located in the top, center of the front panel.
CAUTION – Avoid Warranty Invalidation
Printed circuit assemblies (PCAs) are sensitive to electro-static discharges too small
to be felt by the human nervous system. Damage resulting from failure to use ESD
protection when working with electronic assemblies will void the instrument warranty.
For more information on preventing ESD damage see A Primer on Electro-Static
Discharge;the manual can be downloaded from our website at http://www.teledyne-api.com
under Help Center > Product Manuals in the Special Manuals section.
CAUTION – Avoid Damage to the Instrument
Do not operate the instrument without first removing dust plugs from pneumatic ports.
WARNING!
NEVER DISCONNECT ELECTRONIC CIRCUIT BOARDS, WIRING HARNESSES
OR ELECTRONIC SUBASSEMBLIES WHILE THE UNIT IS UNDER POWER.