Teledyne API T703/T703U Calibrator Operation Manual Operating the Calibrator
57
The GENERATE Step 4.5.2.1.
This step operates and is programmed similarly to the GENERATE > AUTO.
At the end of the programming sequence, the T703/T703U firmware will
automatically insert a DURATION step that needs to be defined.
To insert a GENERATE step into a sequence, press:
INSERT STEP Submenu
SETUP X.X INSERT STEP: GENERATE
PREV NEXT ENTR EXIT
Starting at the INSERT STEPS
Submenu
EXIT discards the new
target concentration
ENTR accepts the new
target concentration
SETUP X.X GENERATE:0.0 PPB O3
0 .0 0 0 PCT O3 ENTR EXIT
SETUP X.X GENERATE:ZERO
ZERO ENTR EXIT
SETUP X.X 3) DURATION : 10.0 MIN
PREV NEXT INS DEL EDIT EXIT
Toggle to switch
between ZERO
AIR and O
3
modes.
SETUP X.X GENERATE:0.0 PPB O3
0 0 0 .0 PPB O3 ENTR EXIT
Toggle these
buttons to set the
target
concentration.
Toggle to scroll
through the
available units of
measure
SETUP X.X DURATION: 1.0 MIN
0 0 1 .0 ENTR EXIT
Toggle these
buttons to set
DURATION of this
step